1. บทนำและผลิตภัณฑ์view
The Apera Instruments EC820 Benchtop Conductivity Meter is a high-precision instrument designed for accurate measurement of Conductivity (EC), Total Dissolved Solids (TDS), Salinity, Resistivity, and temperature. This meter is equipped with advanced features for reliable and stable readings across a wide range of applications.
This manual provides detailed instructions for the proper setup, operation, maintenance, and troubleshooting of your EC820 meter to ensure optimal performance and longevity.

Figure 1: Apera Instruments EC820 Benchtop Conductivity Meter Kit. The image shows the main meter unit, a flexible electrode arm and holder, and the conductivity electrode submerged in a beaker of liquid.
2. ข้อควรระวังเพื่อความปลอดภัย
- Always handle the meter and electrodes with care to prevent damage.
- Do not immerse the main meter unit in water. The unit is IP54 water-resistant, but not submersible.
- ใช้เฉพาะอะแดปเตอร์ที่ให้มาเท่านั้น
- Keep calibration solutions and samples away from children and pets. Follow safety data sheets (SDS) for handling specific chemicals.
- Ensure the meter is placed on a stable, level surface during operation.
- ตัดการเชื่อมต่อไฟฟ้าก่อนทำความสะอาดหรือดำเนินการบำรุงรักษา
3. เนื้อหาบรรจุภัณฑ์
ตรวจสอบว่ามีสินค้าทั้งหมดที่ระบุไว้ด้านล่างอยู่ในแพ็คเกจของคุณ:
- EC820 เครื่องวัดค่าการนำไฟฟ้าแบบตั้งโต๊ะที่มีความแม่นยำ
- 2401T-F Platinum Glass-body Conductivity Electrode
- Flexible Electrode Arm and Holder
- สาย USB
- PC-Link Software Flash Drive
- Power Adaptor (Input: 120V AC 60Hz; Output: 9V 300mA)
- โซลูชันการสอบเทียบ:
- 84 µS/cm (1 vial)
- 1413 µS/cm (1 vial)
- 12.88 mS/cm (1 vial)

Figure 2: The 2401T-F Platinum Glass-body Conductivity Electrode, showing its connection cables.
4. การตั้งค่า
- Unpack the Meter: นำส่วนประกอบทั้งหมดออกจากบรรจุภัณฑ์อย่างระมัดระวัง
- Assemble Electrode Arm: Attach the flexible electrode arm to the meter unit. Ensure it is securely fastened.
- Connect Electrode: Plug the 2401T-F Conductivity Electrode into the designated "COND" port on the back of the meter. Ensure the connection is firm.

Figure 3: Rear panel of the EC820 meter, highlighting the "COND" and "TEMP" ports, along with RS232 and DC 9V power input.
- เชื่อมต่อพลังงาน: Plug the power adapter into the "DC 9V" port on the back of the meter, then connect it to a standard 120V AC 60Hz power outlet.
- เปิดเครื่องครั้งแรก: Press the power button to turn on the meter. The LCD display should illuminate.
5. คู่มือการใช้งาน
5.1. การสอบเทียบ
The EC820 meter features 1 to 4 points auto-calibration. Regular calibration ensures accurate measurements.
- Prepare Solutions: Use the provided 84 µS/cm, 1413 µS/cm, and 12.88 mS/cm calibration solutions. Ensure solutions are at room temperature.
- Rinse Electrode: Rinse the conductivity electrode thoroughly with distilled or deionized water before and after each measurement or calibration. Gently blot dry with a lint-free tissue.
- เข้าสู่โหมดการสอบเทียบ: เมื่อเปิดมิเตอร์แล้ว ให้กดปุ่ม แคล button. The display will show "CAL" and prompt for the first calibration point.
- First Point Calibration: Immerse the electrode into the 1413 µS/cm standard solution. Wait for the reading to stabilize (indicated by a stable icon on the display). The meter will automatically recognize the solution. Press เข้า เพื่อยืนยัน.
- Subsequent Points (Optional): For higher accuracy, proceed with 84 µS/cm and 12.88 mS/cm solutions following the same procedure. The meter will guide you through the process.
- Exit Calibration: After completing desired calibration points, press the มส. button to return to measurement mode.
บันทึก: The meter has a self-diagnosis function to help identify and fix common calibration problems. Refer to the on-screen prompts.
5.2. การวัด
- Rinse Electrode: ล้างอิเล็กโทรดด้วยน้ำกลั่น
- Immerse Electrode: จุ่มอิเล็กโทรดลงในสารละลายample solution, ensuring the sensor is fully submerged.
- Wait for Stability: Allow the reading to stabilize. The stable reading icon will appear on the LCD.
- การอ่านบันทึก: Read the Conductivity (EC), TDS, Salinity, or Resistivity value displayed on the large LCD. Temperature will also be displayed simultaneously.
- เปลี่ยนหน่วย: กดปุ่ม หน่วย button to cycle through different measurement units (e.g., µS/cm, mS/cm, ppm, ppt).
- การจัดเก็บข้อมูล: The meter supports GLP data management with 500 groups of data storage. Use the M+ / R-M buttons to store and recall data.
- การส่งออกข้อมูล: Connect the meter to a PC via the USB cable and use the provided PC-Link software (Windows-based system only) to export data.
6. การบำรุงรักษา
- การทำความสะอาดอิเล็กโทรด:
- Regularly rinse the electrode with distilled or deionized water.
- For stubborn deposits, refer to the electrode's specific cleaning instructions. Generally, a mild detergent solution or a specific electrode cleaning solution can be used, followed by thorough rinsing.
- Do not scratch or damage the platinum black sensor.
- การจัดเก็บอิเล็กโทรด:
- Always store the conductivity electrode in its protective cap filled with a proper electrode storage solution (e.g., 3M KCl solution or Apera Instruments storage solution). Never store dry.
- การทำความสะอาดมิเตอร์: Wipe the main meter unit with a clean, damp ผ้า ห้ามใช้สารทำความสะอาดที่มีฤทธิ์กัดกร่อนหรือตัวทำละลาย
- การอัปเดตซอฟต์แวร์: Check the Apera Instruments website periodically for any software updates for the PC-Link software.

รูปที่ 4: ระยะใกล้ view of the platinum glass-body conductivity electrode tip, which requires careful handling and proper storage.
7 การแก้ไขปัญหา
| ปัญหา | สาเหตุที่เป็นไปได้ | สารละลาย |
|---|---|---|
| มิเตอร์ไม่เปิดใช้งาน | ไม่มีแหล่งจ่ายไฟ; อะแดปเตอร์ชำรุด; การเชื่อมต่อหลวม | Check power adapter connection and outlet. Ensure adapter is functional. |
| การอ่านไม่เสถียร | Dirty electrode; air bubbles on sensor; incorrect calibration; temperature fluctuations. | Clean electrode; gently tap electrode to remove bubbles; recalibrate meter; ensure sampอุณหภูมิคงที่ |
| ข้อผิดพลาดในการสอบเทียบ | Contaminated calibration solution; incorrect solution used; damaged electrode. | Use fresh calibration solutions; ensure correct solution for calibration point; inspect electrode for damage. |
| No data transfer to PC | Incorrect USB connection; software not installed or running; driver issues. | Verify USB cable connection; ensure PC-Link software is installed and open; check device manager for driver issues. |
8. ข้อมูลจำเพาะ
| พารามิเตอร์ | ค่า |
|---|---|
| หมายเลขรุ่น | AI532 |
| พารามิเตอร์การวัด | Conductivity (EC), TDS, Salinity, Resistivity, Temperature |
| ช่วงค่าการนำไฟฟ้า | 0 to 2,000 mS |
| ความแม่นยำ | ±0.5% ฟอซ. |
| จุดสอบเทียบ | 1 to 4 points auto-calibration (84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm) |
| การชดเชยอุณหภูมิ | การชดเชยอุณหภูมิอัตโนมัติ (ATC) |
| การจัดเก็บข้อมูล | 500 groups (GLP compliant) |
| การเชื่อมต่อ | USB for PC data export |
| ประเภทจอภาพ | จอ LCD เรืองแสงขนาดใหญ่ |
| แหล่งจ่ายไฟ | 9V DC, 300mA (120V AC 60Hz adapter included) |
| การป้องกันการไหลเข้า | IP54 กันน้ำได้ |
| การรับรอง | CE, RoHS, ISO 9001:2015 |
| ขนาดแพ็คเกจ | 10.63 x 9.88 x 6.61 นิ้ว |
| น้ำหนัก | 6.5 ปอนด์ |
9. การรับประกันและการสนับสนุน
Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of quality and reliability. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments webเว็บไซต์.
For assistance with troubleshooting, calibration, or general inquiries, please contact Apera Instruments customer service. Contact details can typically be found on their official webเว็บไซต์หรือในบรรจุภัณฑ์ของผลิตภัณฑ์
แหล่งข้อมูลออนไลน์: Apera Instruments Store on Amazon