เอ็มเอสเอ 10178557

คู่มือผู้ใช้เครื่องตรวจจับก๊าซหลายชนิด MSA ALTAIR 4XR

Model: 10178557 | Brand: MSA

การแนะนำ

The MSA ALTAIR 4XR Multigas Detector is a robust, handheld, portable gas monitor designed for critical safety applications. It is capable of simultaneously detecting up to four gases: Lower Explosive Limit (LEL), Oxygen (O2), Hydrogen Sulfide (H2S), and Carbon Monoxide (CO). Engineered for durability, this device features a tough over-molded rubber armor, is MIL-STD-810G drop protected, and boasts an IP68 rating for dust and water resistance, ensuring reliable performance in demanding environments.

MSA ALTAIR 4XR Multigas Detector front view

รูปที่ 1 : ด้านหน้า view of the MSA ALTAIR 4XR Multigas Detector, showing the display and control buttons.

คุณสมบัติหลัก

การตั้งค่าและการสอบเทียบ

Before initial use or after prolonged storage, it is essential to perform a fresh air setup (zero calibration) and a span calibration to ensure accurate readings. Always perform these procedures in a clean, fresh air environment free from contaminants.

อุปกรณ์ที่ต้องมี:

Zero Calibration (Fresh Air Setup):

  1. Power on the ALTAIR 4XR detector by pressing and holding the middle button. Allow it to complete its startup sequence and warm-up procedure.
  2. When prompted for a "Fresh Air Setup" (FAS), select "No" to proceed to normal operating mode. This ensures a high-quality zero calibration.
  3. Once in normal operating mode, press and hold the right button until the display shows "ZERO CAL?".
  4. Press the middle button to confirm "Zero Cal". The unit will enter "SENSOR REFRESH" mode. Ensure the detector is in a clean air environment, free from any gas sources or residual gas.
  5. Allow the unit to complete the zero calibration. The display will indicate when it's complete.

การสอบเทียบช่วง:

  1. Ensure the calibration gas cylinder has the correct gas mixture and is not expired. The standard MSA calibration gas mixture includes: Hydrogen Sulfide (20 ppm), Carbon Monoxide (60 ppm), Pentane Simulant (58% LEL), Oxygen (15.0%), and Nitrogen (Balance).
  2. Attach the 0.25 L/min Regulator to the calibration gas cylinder. Ensure the regulator's valve is open slightly to purge any residual air before fully tightening it onto the cylinder. This prevents moisture from entering the cylinder.
  3. Connect the tubing from the regulator to the calibration adapter.
  4. Attach the calibration adapter to the MSA ALTAIR 4XR detector. The adapter should securely cover the sensor inlets.
  5. From the normal operating mode, press and hold the right button until the display shows "SPAN CAL?".
  6. Press the middle button to confirm "Span Cal". The unit will prompt for gas application.
  7. Immediately turn on the gas flow from the cylinder by opening the regulator valve. The detector will begin to read the gas concentrations.
  8. Monitor the display. The values for each gas (LEL, O2, H2S, CO) should stabilize near the concentrations listed on your calibration gas cylinder. The unit will indicate "SPAN PASS" when calibration is successful.
  9. Once "SPAN PASS" is displayed, turn off the gas flow and remove the calibration adapter and tubing from the detector.
Worker using MSA ALTAIR 4XR Multigas Detector in a confined space

Figure 2: The MSA ALTAIR 4XR Multigas Detector in use, demonstrating its compact size and clear display for field applications.

การดำเนินการ

After successful calibration, the ALTAIR 4XR is ready for use. The large, clear display shows real-time gas concentrations. The unit continuously monitors the environment and will activate audible, visual (LEDs), and vibrating alarms if gas levels exceed predefined thresholds or if a man-down event is detected via MotionAlert.

การซ่อมบำรุง

Regular maintenance is crucial for the longevity and accuracy of your MSA ALTAIR 4XR. Keep the device clean and free from debris. Periodically inspect the sensor inlets for blockages. The unit is compatible with MSA GALAXY GX2 Automated Test System for simplified and automated bump testing and calibration, which also helps in managing data and records.

Sensors are designed for long life but may require replacement over time. The device will provide an "end-of-sensor-life" warning. Replacement MSA XCell Sensors are available and can be easily installed in the plug-and-play slots.

การแก้ไขปัญหา

If the device fails a bump test or calibration, first ensure that the calibration gas cylinder is not expired and contains the correct mixture. Check all connections for leaks. If sensor readings are erratic or consistently off, the sensor may be poisoned or at the end of its life and require replacement. For persistent issues, refer to the comprehensive troubleshooting guide in the full product manual or contact MSA support.

ข้อมูลจำเพาะ

ข้อมูลจำเพาะค่า
ยี่ห้อเอ็มเอสเอ
สไตล์O2, H2S, CO
แหล่งพลังงานใช้พลังงานจากแบตเตอรี่
สีสีดำ
ขนาดสินค้าลึก 1.37 นิ้ว x กว้าง 3 นิ้ว x สูง 4.4 นิ้ว
น้ำหนักสินค้า0.51 ปอนด์ (8.1 ออนซ์)
เตือนได้ยินได้ มองเห็นได้
ความชื้นในการทำงาน95 เปอร์เซ็นต์
คะแนนอุณหภูมิบน50 องศาเซลเซียส
ประเภทเซนเซอร์Catalytic, Electrochemical
จำนวนรายการ1
ผู้ผลิตเอ็มเอสเอ
หมายเลขชิ้นส่วน10178557
หมายเลขรุ่นสินค้า10178557
วันที่พร้อมใช้งานครั้งแรก25 มกราคม 2018

การรับประกันและการสนับสนุน

This product is covered by a 4-year warranty. For more information regarding warranty claims or technical support, please visit the official MSA website or contact their customer service. Ensure to register your product to activate the warranty and receive important updates.

For additional assistance, you may also refer to the comprehensive product manual provided with your device or contact a certified MSA service center.

เอกสารที่เกี่ยวข้อง - 10178557

พรีview MSA ALTAIR 4XR Multigas Detector Operating Manual
Comprehensive operating manual for the MSA ALTAIR 4XR Multigas Detector, covering safety regulations, device description, operation, maintenance, technical data, and ordering information.
พรีview ระบบทดสอบอัตโนมัติ MSA GALAXY GX2 สำหรับเครื่องตรวจจับก๊าซ
ระบบทดสอบอัตโนมัติ MSA GALAXY GX2 นำเสนอการทดสอบและการสอบเทียบที่ใช้งานง่ายและชาญฉลาดสำหรับเครื่องตรวจจับก๊าซชนิดเดียว MSA ALTAIR และ ALTAIR PRO และเครื่องตรวจจับก๊าซหลายชนิด ALTAIR 4XR และ ALTAIR 5X มาพร้อมอินเทอร์เฟซที่ใช้งานง่าย การทดสอบอัตโนมัติ และความสามารถในการจัดการยานพาหนะที่ครอบคลุม
พรีview คู่มือการใช้งานเครื่องตรวจจับก๊าซหลายชนิด MSA ALTAIR 5X และ 5XIR
คู่มือการใช้งานที่ครอบคลุมสำหรับเครื่องตรวจจับก๊าซหลายชนิด MSA ALTAIR 5X และ ALTAIR 5XIR ครอบคลุมถึงข้อบังคับด้านความปลอดภัย การใช้งาน การบำรุงรักษา ข้อมูลจำเพาะทางเทคนิค และการแก้ไขปัญหาสำหรับการตรวจจับก๊าซในสภาพแวดล้อมทางอุตสาหกรรม
พรีview หัววัดปั๊ม MSA ALTAIR: เครื่องตรวจจับก๊าซที่ทนทานสำหรับสภาพแวดล้อมอันตราย
สำรวจ MSA ALTAIR Pump Probe เครื่องตรวจจับก๊าซที่ทนทานและเชื่อถือได้ เรียนรู้เกี่ยวกับคุณสมบัติหลัก ข้อมูลจำเพาะทางเทคนิค การรับรอง และข้อมูลการสั่งซื้อสำหรับการใช้งานด้านความปลอดภัยในอุตสาหกรรม
พรีview คู่มือการใช้งานระบบทดสอบอัตโนมัติ MSA Galaxy GX2
คู่มือการใช้งานสำหรับระบบทดสอบอัตโนมัติ MSA Galaxy GX2 ซึ่งมีรายละเอียดเกี่ยวกับการติดตั้ง การตั้งค่า การใช้งาน การบำรุงรักษา และการแก้ไขปัญหาสำหรับการทดสอบเครื่องตรวจจับก๊าซ
พรีview คู่มือการใช้งานเครื่องตรวจจับก๊าซเดี่ยว MSA ALTAIR Pro - คำแนะนำด้านความปลอดภัยและการใช้งาน
คู่มือการใช้งานนี้ให้คำแนะนำโดยละเอียด แนวทางด้านความปลอดภัย ขั้นตอนการสอบเทียบ และการแก้ไขปัญหาสำหรับเครื่องตรวจจับก๊าซ MSA ALTAIR Pro Single Gas Detector เรียนรู้วิธีใช้และบำรุงรักษาเครื่องตรวจจับก๊าซของคุณอย่างมีประสิทธิภาพ