ข้อมูลด้านความปลอดภัยที่สำคัญ
Please read this manual thoroughly before operating the Elitech PHM-230 pH Pen Meter. Retain this manual for future reference.
- Do not immerse the entire meter in water; only the electrode tip is waterproof.
- Keep the electrode clean and moist. Always replace the protective cap after use.
- Store calibration solutions properly, away from direct sunlight and extreme temperatures.
- Handle the glass electrode with care to prevent damage.
- เก็บให้พ้นจากมือเด็ก
สินค้าเกินview
The Elitech PHM-230 is a professional digital pH pen meter designed for accurate measurement of pH and temperature in various solutions. It features automatic temperature compensation (ATC) and a three-point calibration system.
รวมส่วนประกอบ

Image: The Elitech PHM-230 pH meter, calibration solutions, and batteries neatly arranged in its protective plastic case.
- PHM-230 pH Pen Meter (1 unit)
- Plastic Carrying Case (1 unit)
- Standard Calibration Solutions (3 bottles: pH 4.00, pH 6.86, pH 9.18)
- AAA Batteries (2 units, 1.5V)
- คู่มือผู้ใช้ (1 หน่วย)
คุณสมบัติหลัก
- Precision Calibration System: Three-point push-button calibration with auto-buffer recognition for accurate measurements.
- การชดเชยอุณหภูมิอัตโนมัติ (ATC): Ensures reliable readings in varying temperature environments.
- Advanced Electrode Design: Double-junction structure with enhanced reference gel for extended electrode lifespan.
- จอแสดงผลที่ใช้งานง่าย: Large, dual-line LCD screen shows pH and temperature values, along with calibration status indicators.
- การประยุกต์ใช้ที่กว้าง: Suitable for hydroponics, agriculture, drinking water, aquariums, and industrial solutions.

Image: The Elitech PHM-230 pH meter showing a pH reading of 6.54 and a temperature of 27.1°C.
การตั้งค่า
การติดตั้งแบตเตอรี่
- Unscrew the battery compartment cover located at the top of the meter.
- Insert the two 1.5V AAA batteries, ensuring correct polarity (+/-).
- ใส่ฝาครอบช่องใส่แบตเตอรี่อย่างปลอดภัย
การเตรียมตัวเบื้องต้น
- Remove the protective cap from the electrode tip.
- ล้างอิเล็กโทรดด้วยน้ำกลั่นหรือน้ำปราศจากไอออน
- Gently blot dry with a clean tissue. Do not rub.
การดำเนินการ
การเปิด/ปิดเครื่อง
- กดปุ่ม เปิด/ปิด ปุ่มเพื่อเปิดมิเตอร์
- กดปุ่มค้างไว้ เปิด/ปิด กดปุ่มค้างไว้สองสามวินาทีเพื่อปิดมิเตอร์
การวัด

Image: The Elitech PHM-230 pH meter being used to test the pH of a nutrient solution in a hydroponic system, showing a reading of 3.44 pH and 27.1°C.
- Ensure the meter is clean and calibrated.
- Immerse the electrode tip into the solution to be tested. Ensure the liquid level is above the glass bulb.
- Gently stir the meter and wait for the reading to stabilize on the display.
- Record the pH and temperature values.
- หลังจากทำการวัดเสร็จแล้ว ให้ล้างอิเล็กโทรดด้วยน้ำกลั่นและใส่ฝาครอบป้องกันกลับเข้าที่

Image: The Elitech PHM-230 pH meter showing its measuring range (0-14), pH resolution (0.01), and pH accuracy (±0.05).
การชดเชยอุณหภูมิอัตโนมัติ (ATC)
The PHM-230 automatically adjusts pH readings based on the solution's temperature, ensuring accuracy without manual temperature correction.
Red Light Alarm

Image: The Elitech PHM-230 pH meter showing a red light alarm on its display, indicating that the measured pH is either less than 3.5 or greater than 11.5, signifying a strong acid or strong alkaline environment.
When the pH reading is less than 3.5 or greater than 11.5, the meter will emit a red light alarm. This indicates that the environment is strongly acidic or strongly alkaline.
ฟังก์ชั่นปิดอัตโนมัติ

Image: The Elitech PHM-230 pH meter being used to test a solution, highlighting its 3-point auto calibration, LED backlight for dark environments, and 10-minute auto-off feature to conserve battery life.
The meter is equipped with a 10-minute auto-off function to conserve battery life. If no operation is detected within 10 minutes, the meter will automatically power down. This setting is retained for the next startup.
การสอบเทียบ
Regular calibration ensures the accuracy of your pH meter. The PHM-230 supports 3-point automatic calibration using standard buffer solutions (pH 4.00, pH 6.86, pH 9.18).
Video: An official guide demonstrating the operation and calibration procedure for the Elitech PHM-230 pH Meter.
Calibration Procedure (3-Point)

Image: The Elitech PHM-230 pH meter immersed in three separate beakers containing pH 4.00 (pink), pH 6.86 (green), and pH 9.18 (blue) calibration buffer solutions.
- เปิดเครื่องมิเตอร์
- กดปุ่มค้างไว้ แคล button to enter calibration mode. The display will show 'CAL'.
- Immerse the electrode (glass bulb fully) into the first standard buffer solution (e.g., pH 6.86).
- Wait for the reading to stabilize. The meter will automatically recognize the buffer and save the calibration.
- Rinse the electrode with distilled water for at least 1 minute.
- Repeat steps 3-5 for the second standard buffer solution (e.g., pH 4.00).
- Repeat steps 3-5 for the third standard buffer solution (e.g., pH 9.18).
- After completing all three points, the meter will automatically exit calibration mode and return to measurement mode. Calibration is saved.
การซ่อมบำรุง
Electrode Care and Storage
- Always keep the electrode clean. Rinse with distilled water after each use.
- For long-term storage, ensure the electrode is kept moist. The protective cap should contain a small amount of electrode storage solution or pH 4 buffer solution. Do not store in distilled water.
- If the electrode becomes dry, soak it in electrode storage solution or pH 4 buffer for at least 30 minutes before use.
การทำความสะอาด
- Wipe the meter body with a clean, damp ผ้า ห้ามใช้สารทำความสะอาดที่มีฤทธิ์กัดกร่อนหรือตัวทำละลาย
- If the electrode is contaminated with oil or grease, clean it with a mild detergent solution, then rinse thoroughly with distilled water.
การแก้ไขปัญหา
| ปัญหา | สาเหตุที่เป็นไปได้ | สารละลาย |
|---|---|---|
| การอ่านที่ไม่ถูกต้อง | Uncalibrated meter, dirty electrode, expired buffer solutions, dry electrode. | Calibrate the meter, clean the electrode, use fresh buffer solutions, rehydrate electrode. |
| No display/Meter won't turn on | แบตเตอรี่หมด หรือติดตั้งแบตเตอรี่ไม่ถูกต้อง | Replace batteries, check battery polarity. |
| เวลาตอบสนองช้า | Dry electrode, contaminated electrode. | Rehydrate electrode, clean electrode. |
| การอ่านจะผันผวน | Air bubbles on electrode, unstable sample temperature, electrical interference. | Gently tap meter to remove bubbles, ensure stable sample temperature, move away from electrical sources. |
ข้อมูลจำเพาะ
| คุณสมบัติ | รายละเอียด |
|---|---|
| ขนาดสินค้า | 6.5 x 1.38 x 1.38 นิ้ว |
| น้ำหนักสินค้า | 15.8 ออนซ์ |
| หมายเลขรุ่นสินค้า | พีเอชเอ็ม-230 |
| แบตเตอรี่ | แบตเตอรี่ AAA ขนาด 2 x 1.5V (รวมอยู่ด้วย) |
| ช่วงการวัดค่า pH | ค่า pH 0.00 - 14.00 |
| ความละเอียด pH | 0.01 pH |
| ความแม่นยำของ pH | ±0.05พีเอช |
| ช่วงวัดอุณหภูมิ | 0.0 - 60.0 องศาเซลเซียส (32.0 - 140.0 องศาฟาเรนไฮต์) |
| ความละเอียดของอุณหภูมิ | 0.1 องศาเซลเซียส / 0.1 องศาฟาเรนไฮต์ |
| ความแม่นยำของอุณหภูมิ | ±0.5 °C / ±1.0 °F |
| การชดเชยอุณหภูมิอัตโนมัติ (ATC) | ใช่ |
| การสอบเทียบ | 3-point automatic calibration (pH 4.00, 6.86, 9.18) |
| Auto-Off | 10 นาที |
| ระดับการกันน้ำ | IP67 (electrode tip only) |
ข้อมูลการรับประกัน
For detailed warranty information regarding your Elitech PHM-230 pH Pen Meter, please refer to the official Elitech website or contact Elitech customer support directly. Warranty terms and conditions may vary by region and purchase date.
การสนับสนุนลูกค้า
If you encounter any issues or have questions about your Elitech PHM-230 pH Pen Meter that are not covered in this manual, please contact Elitech customer service for assistance.
- เยี่ยมชมเว็บไซต์อย่างเป็นทางการของ Elitech webเว็บไซต์นี้เป็นแหล่งรวบรวมข้อมูลสนับสนุนและรายละเอียดการติดต่อ
- Refer to the contact information provided on the product packaging or warranty card.