URIT PH-01/PH-01A

URIT PH-01/PH-01A Hemoglobin Test Strips User Manual

For use with URIT Blood Hemoglobin Test Meter PH-01/PH-01A

1. บทนำ

This manual provides essential instructions for the proper use, storage, and maintenance of URIT PH-01/PH-01A Hemoglobin Test Strips. These test strips are designed for use with the URIT Blood Hemoglobin Test Meter PH-01/PH-01A to measure hemoglobin levels in capillary or venous whole blood samples. Please read this manual thoroughly before using the test strips to ensure accurate results and safe operation.

2. สินค้าหมดview

The URIT PH-01/PH-01A Hemoglobin Test Strips are a key component of the URIT hemoglobin testing system. Each package contains 50 test strips, designed for single use.

URIT PH-01 Hemoglobin Test Strips box with two vials, lancets, and capillary tubes

Figure 2.1: Contents of the URIT PH-01 Hemoglobin Test Strips package, including the box, two vials of 25 test strips each, and accompanying lancets and capillary tubes.

2.1 คุณสมบัติหลัก

  • ปริมาณ: 50 test strips per package.
  • ความเข้ากันได้: Specifically designed for URIT Blood Hemoglobin Test Meter PH-01/PH-01A.
  • ความสะดวกในการใช้งาน: Simple procedure for blood sampแอปพลิเคชัน
  • ความแม่นยำ: Provides reliable hemoglobin measurements when used correctly.
  • ผลลัพธ์ที่รวดเร็ว: Designed for quick measurement times.
Graphic illustrating key features: Professional, Easy to Use, Accurate, 20 Seconds

Figure 2.2: Visual representation of the key attributes of the URIT Hemoglobin Test Strips, highlighting their professional quality, ease of use, accuracy, and quick results within 20 seconds.

3. ข้อมูลด้านความปลอดภัยและข้อควรระวัง

Always follow standard laboratory safety procedures when handling blood samples. Dispose of used test strips, lancets, and any other contaminated materials according to local regulations for biohazardous waste.

การปฏิเสธความรับผิดทางกฎหมาย:

คำกล่าวอ้างเกี่ยวกับอาหารเสริมยังไม่ได้รับการประเมินจาก FDA และไม่ได้มีจุดประสงค์เพื่อวินิจฉัย รักษา รักษาหาย หรือป้องกันโรคหรือภาวะสุขภาพใดๆ

  • สำหรับ ในหลอดทดลอง diagnostic use only. Do not ingest.
  • เก็บให้พ้นจากมือเด็ก
  • Store test strips in their original vial with the cap tightly closed to protect from moisture and light.
  • ห้ามใช้แถบตรวจที่หมดอายุแล้ว โปรดตรวจสอบวันหมดอายุบนขวดบรรจุ
  • Do not use test strips if the vial is damaged or has been left open.
  • Consult a healthcare professional for interpretation of results and medical advice.
Diagram showing risks associated with high and low hemoglobin levels, including kidney disease, liver disease, anemia, chronic lung disease, dehydration, and heart failure.

Figure 3.1: Illustrative diagram outlining potential health risks associated with both low and elevated hemoglobin levels. This information is for general awareness and not for self-diagnosis.

4. การตั้งค่าและการเตรียมการ

4.1 Materials Required (Not included with test strips)

  • URIT Blood Hemoglobin Test Meter PH-01/PH-01A
  • Lancet device and sterile lancets
  • ทิชชู่เปียกแอลกอฮอล์
  • สำลีหรือผ้ากอซ
  • ภาชนะกำจัดสารอันตรายทางชีวภาพ
Image showing a vial of URIT PH-01 Hemoglobin Test Strips, capillary tubes, and lancets, indicating items included in the package.

Figure 4.1: Components typically included with the URIT PH-01 Hemoglobin Test Strips package: test strip vial, capillary tubes, and lancets.

4.2 Preparing for a Test

  1. Ensure the URIT Hemoglobin Test Meter is clean and functioning correctly.
  2. Wash your hands thoroughly with soap and water, then dry them completely. This helps prevent contamination.
  3. Prepare the lancing device with a new, sterile lancet.
  4. Open the test strip vial and remove one test strip. Immediately close the vial tightly to protect the remaining strips.
  5. Insert the code card (if a new vial) into the meter to calibrate it. Remove the code card after calibration. (Note: The same bottle of test strips does not need to repeat the settings.)
  6. Insert the test strip into the meter in the direction indicated by the arrows on the strip. The meter should turn on automatically or display a prompt for blood application.
Four-step diagram for blood collection: 1. Wash hands, 2. Prepare lancing device, 3. Screw down cap, 4. Adjust lancing depth.

Figure 4.2: Detailed steps for preparing for blood collection, emphasizing hygiene and proper setup of the lancing device.

5. Operating Instructions (Performing a Test)

Follow these steps carefully to perform a hemoglobin test using the URIT PH-01/PH-01A Hemoglobin Test Strips and Meter.

Six-step quickstart guide for PH-01: 1. Power on meter, 2. Insert code card, 3. Insert test strip, 4. Draw blood sample, 5. Apply blood to strip, 6. Read results.

Figure 5.1: Quickstart guide illustrating the six main steps for performing a hemoglobin test with the URIT PH-01 meter and test strips.

  1. Power On Meter: Press and hold the power button on the URIT Hemoglobin Test Meter.
  2. Insert Code Card (if new vial): If using a new vial of test strips, insert the code card provided with the strips into the meter. The meter will read the code. Remove the code card once calibration is complete. (This step is not required for subsequent tests from the same vial).
  3. ใส่แถบทดสอบ: Turn on the hemoglobin meter and insert a fresh hemoglobin test strip into the designated slot.
  4. รับเลือด Sampเลอ: Use the prepared lancing device to prick the side of your fingertip (or other approved site) to obtain a blood drop. Gently massage the finger to form a sufficient blood drop.
  5. ใช้เลือด Sampเลอ: When the flashing blood drop symbol appears on the meter screen, press the tip of the test strip to the blood sample. The blood will automatically be drawn into the sample entry port of the test strip. Ensure the sample completely fills the reaction area.
  6. อ่านผลลัพธ์: The meter will begin the analysis. Read the test results displayed on the meter screen within approximately 20 seconds.
Close-up of URIT PH-01 meter with a test strip inserted and a blood sample being applied, demonstrating easy operation.

Figure 5.2: Illustration of the easy operation, showing a small amount of blood sample being applied to the test strip inserted into the URIT PH-01 meter.

5.1 Post-Test Procedure

  • Apply gentle pressure with a cotton ball or gauze to the lanced site until bleeding stops.
  • Carefully remove the used test strip from the meter.
  • Dispose of the used test strip and lancet immediately into a proper biohazard sharps container.
  • Record your results if necessary.

6. การบำรุงรักษาและการจัดเก็บ

6.1 การเก็บรักษาแถบทดสอบ

  • Store URIT Hemoglobin Test Strips in their original, tightly capped vial.
  • Keep the vial in a cool, dry place, away from direct sunlight and heat.
  • Do not refrigerate or freeze the test strips.
  • Strips are stable for 3 months after first opening the vial, provided they are stored correctly. Note the opening date on the vial.
  • Do not use test strips beyond the expiration date printed on the vial.

6.2 Meter Maintenance

  • Clean the meter regularly with a soft, damp ผ้า. อย่าใช้สารเคมีที่รุนแรงหรือน้ำยาทำความสะอาดที่มีฤทธิ์กัดกร่อน
  • Ensure the meter's test strip port is free of dust or debris.
  • Store the meter in a protective case when not in use.

7 การแก้ไขปัญหา

If you encounter issues during testing, refer to the following common problems and solutions. For persistent issues, contact URIT customer support.

ปัญหาสาเหตุที่เป็นไปได้สารละลาย
มิเตอร์ไม่ทำงานแบตเตอรี่อ่อนหรือใส่แถบตรวจไม่ถูกต้องCheck meter battery and replace if necessary. Ensure test strip is inserted correctly.
ข้อความแสดงข้อผิดพลาดบนมิเตอร์Expired strip, insufficient blood sample, or strip damage.Use a new, unexpired test strip. Ensure adequate blood sample is applied. Check strip for damage before use.
เลือด sampไม่ได้วาดเป็นแถบInsufficient blood volume or incorrect application.Ensure a sufficiently large blood drop. Touch the tip of the strip to the blood drop, allowing it to wick in naturally. Do not smear.
ผลลัพธ์ที่ไม่สอดคล้องกันหรือไม่คาดคิดImproper storage, expired strips, contamination, or meter malfunction.Verify proper strip storage and expiration date. Ensure hands and lancing site are clean. Perform a control solution test if available. Contact customer support if issues persist.

8. ข้อมูลจำเพาะ

  • ชื่อสินค้า : URIT Hemoglobin Test Strips
  • ความเข้ากันได้ของรุ่น: URIT Blood Hemoglobin Test Meter PH-01/PH-01A
  • Sampประเภทไฟล์: Capillary or venous whole blood
  • เวลาทดสอบ: ประมาณ 20 วินาที
  • อายุการเก็บรักษา: Refer to expiration date on vial. Stable for 3 months after opening.
  • เนื้อหาแพ็คเกจ: 50 Test Strips (2 vials of 25 strips each), Code Card, Capillary Tubes (50), Lancets (50).
  • ผู้ผลิต: ยูริท
  • ASIN: B0BHVBFBDL

9. การรับประกันและการสนับสนุนลูกค้า

For warranty information regarding the URIT Hemoglobin Test Strips or the URIT Blood Hemoglobin Test Meter, please refer to the documentation provided with your meter or contact URIT customer support directly. Test strips are consumable items and typically do not carry a separate warranty beyond being free from manufacturing defects upon purchase.

9.1 ข้อมูลการติดต่อ

For technical assistance, product inquiries, or to report issues, please contact URIT Official customer support:

  • ชื่อผู้ขาย: URIT Official
  • ลิงก์ผู้ขายบน Amazon: URIT Official Seller Page
  • ร้านค้าแบรนด์: URIT Amazon Store
  • ผู้ผลิต: ยูริท

Please have your product model number (PH-01/PH-01A) and ASIN (B0BHVBFBDL) ready when contacting support.

© 2024 URIT. All rights reserved.

คู่มือนี้จัดทำขึ้นเพื่อเป็นข้อมูลเท่านั้น และไม่ควรใช้แทนคำแนะนำทางการแพทย์จากผู้เชี่ยวชาญ

Ask a question about this manual